Invivo2 500

  • Specifications

    Workstation Fabrication

    • Solvent Bonded Acrylic

    Bench Space Requirements

    • Height (excluding stand): 750 mm
    • Width (side to side, excludes Gas Mixer): 1420 mm
    • Depth: 720 mm

    Internal Workstation Dimensions

    • Height: 480 mm
    • Width: 800 mm
    • Depth: 500 mm

    Workstation Weight

    • Weight approx: 140kg

    Gas Supply

    • Up to 4 Separate Cylinders: H2/N2; CO2; N2; Air

    Alarms

    • Temperature: Visual
    • Humidity: Visual
    • Gas Low Pressure: Visual and Audible

    Interlock Dimension

    • Height: 300 mm
    • Width: 320 mm
    • Depth: 300 mm

    Interlock Cycle Time

    • 5 minutes

    Atmosphere Control

    • Temperature: Ambient +5°C - 45°C
    • Humidity Control: Ambient-85% RH
    • O2 Control: 0.0%-20.9%
    • CO2 Control: 0.0%-30.0%
    • Activated Carbon Filtration System
    Capacity Checker

    Use the Capacity Checker to compare how many Plates/Flasks/Dishes fit into each Workstation

  • Accessories

    Standard Accessories

    Workstation

    • Gas Mixer Q
    • Removable Front Panel
    • Internal Mains Socket
    • Single Plate Entry System
    • Direct-hand Access using Ezee Sleeve™ System
    • Detox Sachet
    • Energy Saving Flourescent Lamp
    • Internal Halogen Spot Lamp

    Gas Control Software

    • USB Communications Port
    • 7 Days Continuous Storage of Event Log Data
    • One Touch Calibration (O2)
    • On-Screen Fault Assistance
    • Cycle Programming

    Optional Accessories

    • Cable Gland Port to allow the passage of cables through the workstation wall without altering internal conditions
    • Vacuum Line Port to allow the attachment of a vacuum line used to remove waste media from plates and dishes without removing them from the hypoxic conditions inside the workstation
    • Gas Sample Port to enable samples of the Workstation atmosphere to be checked
    • Ultrasonic Humidity Control for improved cell conditions
    • Stand to support workstation where bench space is not available
    • Palladium Catalyst if anaerobic environment is required
    • Anaerobic Indicator Strips to check anaerobic state of internal environment if required
    • Gas Tank Regulators
    • Gas Tank Filter Modules
    • Data Logging Connection allows the user to monitor the temperature of the workstation via a PC
    • Gas Level Verification Meter designed to monitor CO2, and O2 inside Ruskinn Workstations, works under high humidity.
    Find accessories for this product

    Find accessories for Ruskinn Workstations

  • Downloads