Distributor Support: Login | Register

Sci-tive N Workstation

Product Image:

  • Click image to enlarge

Specifications:

  • External Workstation Dimensions

  • Height (excluding stand): 1024 mm
  • Width (side to side): 1650 mm
  • Depth: 760 mm
  • (additional 150mm required behind workstation for piping)
  • Internal Workstation Dimensions

  • Height: 590 mm
  • Width: 1174 mm
  • Depth: 570 mm
  • Workstation Weight

  • Weight approx: 230kg
  • Interlock Dimension

  • Height: 210 mm
  • Width: 270 mm
  • Depth: 200 mm
  • Interlock Capacity

  • 11.3 L
  • Atmosphere Control

  • Temperature: Ambient + 5°C- 45°C
  • Humidity Control: Ambient-85%
  • O2 Control: 0.0%-20.9%
  • CO2 Control: 0.0%-30.0%
  • Integrated HEPA Filtration System
  • Activated Carbon Filtration System
  • Alarms

  • Temperature: Visual and Audible
  • Humidity: Visual
  • Gas Low Pressure: Visual and Audible

Downloads/Links:

Sci-tive has been designed to mimic in-vivo conditions providing a continuous cell culture environment which eliminates cellular stress linked to variations in temperature, pH and oxidation yielding better stem cell lines.

Features and Benefits

Complete 'Lab in a Box'

Reduces need for laminar air flow cabinets, stereo microscopes, CO2 incubators

 

Stress-Free Cell Manipulation

Now Stem Cell Biologists can work under stable O2, CO2, pH, temperature and humidity for derivation, manipulation and culturing steps

 

Direct-Hand Access into Workstation

Using sterilised and user friendly cuff & sleeve system (Ezee Sleeve™)

 

Aids Compliance with EU TCD (2004/23/EU)

VOC filtration and HEPA filtration create a biologically safe working environment

 

CE (1293) Approved (MDD 93/42/EEC)

Compliant with European EMC and electrical safety tests

Standard Accessories

Workstation

  • Stand
  • 2 Racks with Perforated Shelves to Optimise Homogeneity
  • Direct-hand Access using Ezee Sleeve™ and Ezee Pleez System
  • Removable Front with UV Filtration Screen
  • Waste trays
  • Gas tight waste port

Gas Control Software

  • USB Communications Port
  • 7 Days Continuous Storage of Event Log Data
  • One Touch Calibration (O2)
  • On-Screen Fault Assistance

Optional Accessories

  • Cable Gland Port to allow the passage of cables through the workstation wall without altering internal conditions
  • Vacuum Line Port to allow the attachment of a vacuum line used to remove waste media from plates and dishes without removing them from the hypoxic conditions inside the workstation
  • Gas Sample Port to enable samples of the Workstation atmosphere to be checked
  • Palladium Catalyst if anaerobic environment if required
  • Anaerobic Indicator Strips to check anaerobic state of internal environment if required
  • Gas Tank Regulators
  • Gas Tank Filter Modules
  • Power Failure Backup System maintains the hypoxic environment inside workstation if there is a mains power failure
  • Vibration Isolation Platform designed to minimise vibration originating from the Workstation or Laboratory Floor; can be used for vibration free incubation of microtitre plates, dishes or flasks
  • Adjustable incubation tray so that any micro-titre plates, flasks or dishes are kept absolutely level during incubation
  • Adjustable chair
  • Remote Monitoring Module allows users to be alerted by text message when gas pressure of O2 and gas pressure of CO2, temperature or humidity levels are out of specification
  • Choice of "Free Flow" incubation racks
    1: 150mm deep x 220mm tall x 460mm wide
    2: 150mm deep x 320mm tall x 290mm wide
  • Gas Level Verification Meter (Geotech 100) designed to monitor CO2, and O2 inside Ruskinn Workstations, works under high humidity